Rice Lake TS Series Tuning Fork Stainless Steel Electronic Balance Operation Manual • To ensure safe and proper use o
6 RL-TS Series Operation Manual 1BHow to Read this Manual This manual comprises the following chapters: Chapter 1 How to Begin This chapter gives i
96 RL-TS Series Operation Manual Minimum Display by Unit of Measurement Model Unit_of measuremant TS- 220 TS- 320 TS- 620 TS- 820 TS- 2200 TS-
RL-TS Series Operation Manual 97 Model Unit_of measuremant TS- 15K g 15000 1 kg 15 0.001 (ct) 75000 5 (oz) 520 0.05 (lb) 33 0.005 (o
98 RL-TS Series Operation Manual 40BAppendix 5 Conversion Table Units Unit Gram carat Ounce Pound troy ounce Penny Weight 1 g 1 5 0.03527 0.0
RL-TS Series Operation Manual 99 TS Series Limited Warranty Rice Lake Weighing Systems (RLWS) warrants that all RLWS equipment and systems manu
RL-TS Series Operation Manual 7 1.0 How to Begin This chapter describes the operations required before using the balance, starting from assemb
8 RL-TS Series Operation Manual 3B1. 1 Checking Supplied Items The following items are contained in the box. In the unlikely event of problems su
RL-TS Series Operation Manual 9 4B1. 2 Names and Functions of Component Parts Pan Level Indicates the inclination of the balance. Display O
10 RL-TS Series Operation Manual - When not using the balance, be sure to attach the AC adapter jack cap, connector cover, and the rubber cap and
RL-TS Series Operation Manual 11 1. 3 Workings of Operation Keys The operation keys are provided on the front of the main unit. Use these key
12 RL-TS Series Operation Manual 6B1. 4 How to See Displayed Signs Each of the signs displayed on the front of the main unit has the following mean
RL-TS Series Operation Manual 13 Lights up when sum totals are displayed (shared use together with other readout units) when the addition func
14 RL-TS Series Operation Manual 7B1. 5 Assembling and Installing the Balance Assembling the balance Assemble the main unit with the following step
RL-TS Series Operation Manual 15 Use the balance in a horizontal position. By using the level provided on the front of the main unit and the a
16 RL-TS Series Operation Manual
RL-TS Series Operation Manual 17 2.0 Basic Operation This chapter describes how to use basic measuring functions that are used daily. This cha
18 RL-TS Series Operation Manual 8B2.1 Powering On/Off the Balance and Checking Operation Turning the power of the balance on and off 1 Turn the p
RL-TS Series Operation Manual 19 9B2.2 Weighing by Placing a Sample in a Container (Tare) When measuring weight with the sample in a container
20 RL-TS Series Operation Manual 10B2.3 Weighing an Added Sample Place an additional sample. Only the added weight is measured. 1 Place the sample
RL-TS Series Operation Manual 21 11B2.4 Displaying the Sum of the Container and the Sample The sum weight of the sample and the container is d
22 RL-TS Series Operation Manual 12B2.5 Function Setting Basics To set the settings of the balance, use the function keys. This section describes th
RL-TS Series Operation Manual 23 For the items and settings settable by the function capability, refer to “Appendix 1: Function Setting List”
RL-TS Series Operation Manual 24 3.0 Various Measuring Methods In addition to weighing, the balance has the following measuring modes: parts c
RL-TS Series Operation Manual 25 13B3.1 Weighing (Weighing Machine) By default, the balance is set to “weighing machine” mode. To return to we
RL-TS Series Operation Manual i Contents Contents ... i Safety Precautions ...
26 RL-TS Series Operation Manual 14B3.2 Counting Parts Count The balance saves sample weight (unit weight) using the automatic memory update method
RL-TS Series Operation Manual 27 7 Put additional samples. Put additional samples. The number of additional samples is up to three times the
28 RL-TS Series Operation Manual 15B3.3 Measuring Percentage With respect to the reference sample weight, the weight of a sample is shown in percent
RL-TS Series Operation Manual 29 45BSetting a reference weight by entering a value 1 Set to the function setting mode. (Refer to “Section 2-5:
30 RL-TS Series Operation Manual 7 Weigh. The display indicates the percentage of the sample with respect to the reference sample weight. Pressing
RL-TS Series Operation Manual 31 16B3.4 Obtaining Weight Multiplied by Coefficient Measured weight is multiplied by a set coefficient, and the
32 RL-TS Series Operation Manual 6 Save the coefficient. Press the [Target] key. 7 Weigh. The weight of the sample is multiplied by the coeffic
RL-TS Series Operation Manual 33 3.5 Measuring Specific Gravity The specific gravity of a sample is measured using the underfloor weigher funct
34 RL-TS Series Operation Manual 48BMeasurement 1 Set to the function setting mode. (Refer to “Section 2-5: Function Setting Basics (P. X22X).) Pre
RL-TS Series Operation Manual 35 6 Save the water temperature. Press the [Target] key. 7 Hang only the cage to zero the weight readout.
ii RL-TS Series Operation Manual 6.0 Input/Output to/from External Devices ... 64 29B6.1 Outputting to a Printer ...
36 RL-TS Series Operation Manual 49BUsing a liquid medium other than water 1 Set to the function setting mode. (Refer to “Section 2-5: Function Sett
RL-TS Series Operation Manual 37
38 RL-TS Series Operation Manual 18B3.6 Weighing an Animal The balance can accurately weigh animals and other samples that move during measurement.
RL-TS Series Operation Manual 39 - If the animal moves too much, the value may not be held. - The measurement unit used in animal weighing i
40 RL-TS Series Operation Manual 3.7 Adding Multiple Measurements Multiple samples are weighed consecutively and the sum is displayed. The weighing
RL-TS Series Operation Manual 41 52BWeighing with addition function 1 Place the first sample. After an asterisk (*) is displayed, press t
42 RL-TS Series Operation Manual 20B3.8 Judging “Above” and “Below” (Limit Function) By setting values in the balance, you can judge whether measur
RL-TS Series Operation Manual 43 71B Judgment criteria and limit value setting A limit value can be judged with any of the following criteri
44 RL-TS Series Operation Manual 53BLimit function setting First, set the limit function. Then, set limit values. 1 Set to the function setting mode
RL-TS Series Operation Manual 45 7 Save the setting. Press the [Target] key. Then, set the values for judgment (limit values) to perform measu
RL-TS Series Operation Manual 1 Safety Precautions Before using the balance, please read this manual carefully. The nature of dangers and damage
46 RL-TS Series Operation Manual 74B Judging by absolute values using the method of entering values Set upper and lower limits by entering values
RL-TS Series Operation Manual 47 55BJudging by deviation values The function setting for deviation value judgment is “24.Tp.2.” 75B Judging
48 RL-TS Series Operation Manual 76B Judging by deviation values using the method of entering values Set a reference, upper limit, and lower limit
RL-TS Series Operation Manual 49 5 Weigh. Place a judgment target sample on the pan. The OK, LO, or HI result judged from the upper and lower
RL-TS Series Operation Manual 50 4.0 Adjusting the Scale Calibrate the balance using weights. This chapter includes: How to Calibrate Testing
RL-TS Series Operation Manual 51 21B4.1 How to Calibrate To calibrate a balance is called span adjustment. Be sure to perform span adjustment f
52 RL-TS Series Operation Manual 6 Place the weight on the pan. After zero-point adjustment finishes and the display changes to “on F.S.,” place th
RL-TS Series Operation Manual 53 22B4.2 Testing the Balance The amount of offset with respect to the reference weight can be checked. This is c
54 RL-TS Series Operation Manual 5.0 Function Setting Set the functions of the balance. This chapter includes: Using Two Expression Units by Switc
RL-TS Series Operation Manual 55 23B5-1 Using Two Expression Units by Switching Them You can set two units (unit A and unit B) and switch betw
2 RL-TS Series Operation Manual Danger Do not disassemble or modify the unit. Doing so could cause accidents such as injury, electrical shock, an
56 RL-TS Series Operation Manual 5.2 Minimum Readability Setting Use this function to set the minimum readability. The larger the minimum readabilit
RL-TS Series Operation Manual 57 25B5.3 Saving Container (Tare) Weight Use this function to set a tare range when the balance is powered on usi
58 RL-TS Series Operation Manual 26B5.4 Power Setting 56BAuto Power Off This function, which is only available when the balance is operated by bat
RL-TS Series Operation Manual 59 57BAuto Backlight Off This function automatically turns off the backlight if the balance is left untouched in
60 RL-TS Series Operation Manual 5.5 ID No. Setting Results of successfully completed span adjustments and span tests can be printed by ISO/GLP/GMP
RL-TS Series Operation Manual 61 5 Save the ID number. Press the [Target] key. Display changes to “ ” Pressing the [Target] key ag
62 RL-TS Series Operation Manual 28B5.6 Improving the Stability of the Balance When the balance is stable, “” is lit in the upper left of the disp
RL-TS Series Operation Manual 63 4 Save the setting. Press the [Target] key. The balance goes back to normal display.
RL-TS Series Operation Manual 64 6.0 Input/Output to/from External Devices Balance data can be output to a printer and be input and output to/f
RL-TS Series Operation Manual 65 29B6.1 Outputting to a Printer By connecting a printer to the balance, span adjustment and span test results
RL-TS Series Operation Manual 3 When the balance is not waterproof and dustproof, do not expose the balance to rain, water, dust, or such ot
66 RL-TS Series Operation Manual 5 Perform span adjustment or span test. After span adjustment or test is successfully completed, the printer execu
RL-TS Series Operation Manual 67 30B6.2 Connecting to External Devices via RS-232C Interface The balance inputs and outputs from and to externa
68 RL-TS Series Operation Manual You can set a tare range or adjust the zero-point from an external device by connecting a contact or a transistor
RL-TS Series Operation Manual 69 63BInterface specifications Transmission system Serial transmission, Start-stop synchronization Transmission
70 RL-TS Series Operation Manual 31B6.3 Communication Data and Commands The RS-232C interface exchanges data with external devices as follows: 6
RL-TS Series Operation Manual 71 77B Data format - 6-digit numeric format Consists of 14 characters including terminators (CR = 0DH/LF = 0AH
72 RL-TS Series Operation Manual * -If headed with no data, the numeric value is headed by 0 (30H) by factory default setting. Using the function se
RL-TS Series Operation Manual 73 [S2] (one character) Indicates the status. S2 Code Description S 53H Data stable *1 U 55H Data unstable
74 RL-TS Series Operation Manual 65BInput commands An input command is to control the balance from an external device. The following four input com
RL-TS Series Operation Manual 75 Command format (1) Tare range (zero-point adjustment) command C1 C2 Code (C1) Code (C2) Description V
4 RL-TS Series Operation Manual Do not use the balance in the following places: Location where it may be subject to air from an air-conditioning
76 RL-TS Series Operation Manual - When either an O8 or O9 command is executed, it returns to the state of “O0”. (3) Set measurement mode command
RL-TS Series Operation Manual 77 You can select the response format of either the [“A00”/“Exx”] format or the [ACK/NAK] format. In the explanat
78 RL-TS Series Operation Manual 7. 0 Troubleshooting This chapter describes troubleshooting including error messages and remedies. This chapter
RL-TS Series Operation Manual 79 32B7. 1 Error Messages Message Cause Remedy - The weight of the sample is over the weighing capacity.
80 RL-TS Series Operation Manual Message Cause Remedy The balance is influenced by static electricity or noise. - Unplug the AC adapter from the
RL-TS Series Operation Manual 81 7. 2 Troubleshooting Problem Cause Remedy Nothing is displayed even when the balance is powered on. The AC
82 RL-TS Series Operation Manual Problem Cause Remedy The weight used for calibration is slightly different in mass from the weight used for check
RL-TS Series Operation Manual 83 Problem Cause Remedy No outputs Intended output function settings are not established. Make the balance co
84 RL-TS Series Operation Manual 7.3 Initializing The settings of the balance can be initialized with the following steps: 1 Set to the function 2
RL-TS Series Operation Manual 85 35B7. 4 Maintenance When taking care of the balance, be careful of the following: 67BFor heavy dirt If the b
RL-TS Series Operation Manual 5 Respect the following: Be sure to calibrate the balance after installation or relocation. Measurement values
RL-TS Series Operation Manual 86 Appendixes The appendixes provide data including each type of data for the balance. This chapter includes: Fu
RL-TS Series Operation Manual 87 Appendix 1 Function Setting List Additional Animal weighing selected “Cumulate function” or “Cumulate fu
88 RL-TS Series Operation Manual
RL-TS Series Operation Manual 89
90 RL-TS Series Operation Manual 37B
RL-TS Series Operation Manual 91 Appendix 2 Measurement Mode List In each weighing mode, pressing the [Function] key can toggle the function d
92 RL-TS Series Operation Manual Appendix 3 Printing in Compliance with ISO/GLP/GMP Span adjustment with external weight Span t
RL-TS Series Operation Manual 93 Measurement data: Header Measurement data: Footer The date and time shown in p
94 RL-TS Series Operation Manual 39BAppendix 4 Specifications 83B Basic specifications Model Gram (g) Parts counting Percentage Dimension of
RL-TS Series Operation Manual 95 Output RS-232C compliant output is equipped as standard. D-SUB9P male (RS-232C output, port for external tare