Rice-lake DWM-IV Weighmeter Series 9 Setup Manual Instrukcja Użytkownika Strona 12

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 35
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 11
10
Setup/Configuration Mode Access
There are two ways to access the Setup/Configuration Mode:
1) Using the Calibration Jumper (JP1)
i) Turn off the instrument and insert Jumper 1 (JP1).
ii) Turn the instrument on, the display will show the Audit Trail then
prompt for password entry [P 0].
iii) Enter the password “2001” and press the [Enter] key.
2) Return Jumper 1 (JP1) to its original position when setup/calibration is
completed.
3) “Hot Key” entry if enabled (C1.2 is set to Enable.)
i) Press the [
G/N
] and [
Enter
] keys together while in normal operation
mode
ii) The display will show the Audit Trail then prompt for password entry
[P 0].
iii) Enter the setup/calibration password and press the [Enter] key.
Configuration/Calibration Mode
Category Selection:
The setup mode is arranged into major categories and their individual
selections. Upon entering the setup mode the display will prompt ‘C1’
indicating category ‘1’. Using the ‘Print’ and ‘G/N’ keys scroll through the setup
categories until the desired category is displayed. When the desired category
is displayed press the ‘Enter’ key to begin setup. Pressing ‘Zero’ while in the
category menu will prompt ‘END’, pressing ‘Enter’ at this prompt exits the
setup menu.
Setup Selection:
After entering the desired category use the ‘Print’ and ‘G/N’ keys to toggle
setup choices or alter setup values. Press ‘Enter’ to accept when the desired
value is displayed to save the value and advance to the next selection.
Note: Features requiring the Group B Function keys will not be available
when configuring a standard Series 9.
Przeglądanie stron 11
1 2 ... 7 8 9 10 11 12 13 14 15 16 17 ... 34 35

Komentarze do niniejszej Instrukcji

Brak uwag