16
Configuration
Step
Values
Set
Description
C6 Calibration
REF Calibration Mode
RC
Standard Calibration (Zero Calibration and Span Calibration)
RS
Span Calibration Only
C6.1
RO
Zero Calibration Only
PCAL Zero Calibration
Display will show
PCAL
about 1 sec then “
R XXXX.XX
”
C6.2
Insure the scale is unloaded and press ‘Enter’
CAL Span Calibration
Display will show “CAL” about 1 sec then “N XXXX.XX”
Key in weight value to be used for Span Calibration (Optional Series 9)
Scroll in the weight value to be used for Span Calibration (Std. Series 9)
Place the Weight on the scale and p ress ‘Enter’
C6.3
Press Enter key to save the Calibration when “End” is displayed
C7
Calibration Tune
C7.1 TUN Tune Calibration (Adjust Span)
0~255
1 = 0.1 times the current value of the one division
C7.2 DIR
Tune Direction
UP Adjust span upward
DOWN
Adjust span downward
(Optional Series 9 Unit Only)
To Enter the Custom Print Format Procedure
With the
C1
category displayed press ‘Shift’ then ‘0’
C10
Custom Print Formats (See details in the Customizing Print Format
section)
‘P1 CUST’
C10.1
Edit the Custom Print Format
‘P2 CONT’
C10.2
Edit the Custom Continuous Print Format
‘FMT END’
C10.3
Quit Print Format